| Electronics
|
|
Symptom/Message |
Probable
Cause |
Suggested
Remedy |
| Main
power supply button doesn't light up. |
The AC power cord is unplugged
from the wall socket. |
Plug the power
cord into the wall socket correctly. |
| Computer
power LED is not lit, or the computer doesn't work. |
1. The AC power cord is unplugged
from the computer.
2. The computer's rear power button is switched off
3. The system's fuse has burnt out. |
1. Plug the
power cord into the computer socket (as described in the GI Mentor
Installation Guide).
2. Switch on the computer's rear power button.
3. Contact Simbionix customer support. |
| The
mouse/keyboard doesn't work. |
The mouse/keyboard is not plugged
in. |
Plug in the appropriate
cords (as described in the GI Mentor Installation Guide). |
| Software |
| |
Probable
Cause |
Suggested
Remedy |
| The
computer is turned on, yet I can't see the Login screen. |
Components are not plugged in correctly. |
Shut down the computer,
turn off the main power supply, plug in all components securely, turn
on the main power supply, and restart the computer. |
| I
received the message: "The password is incorrect". |
The password was incorrectly typed. |
Check the password,
and enter it again using capital letters and lower case letters, as
required. |
| The
sound doesn't work |
1. The sound cord is unplugged.
2. The sound is muted. |
1. Plug in
the sound cable (as described in the GI Mentor Installation Guide).
2. Press the "Mute" button on the screen. |
| Simulator |
| |
Probable
Cause |
Suggested
Remedy |
| "I
want to start a case, the scope is inside the mannequin, yet nothing
is happening. The procedure doesn't start
" |
1. One of the Gastroscopy cases
was selected when the mannequin was in its Colonoscopy position,
or vice versa.
2. The scope is not correctly plugged in.
3. The scope wasn't re-calibrated. |
1. Turn the
mannequin or select the appropriate case.
2. Push the scope tightly into its computer's connector.
3. Contact Simbionix customer support. |
| "I
want to start a cyberscopy exercise, the scope is inside the mannequin,
yet nothing is happening. The exercise doesn't start
" |
The mannequin was in its Gastroscopy
position and the scope was inserted into the mouth |
Turn the mannequin
to its Colonoscopy position and try again. |
| "The
Map, Air and other Indicators don't work." |
As a Trainee, you may have limited
options defined by the Instructor. |
Refer to "Enabling/
Disabling Instructional Aids" in the Instructor's Manual. |
| I
received the message:"Get matrix failed". |
The cable connectors are not correctly
plugged in. |
Shut down the computer,
turn off the main power supply, plug in all components securely, turn
on the main power supply, and restart the computer. |
| Up
|
|
|